Page 58
Section 4.2 Using the System Configuration Tool (SUIF)
4.
On the Utilities desktop, next to Service, press LAUNCH.
The Service Home tab is displayed. The Home tab contains summary information including
site identification, connectivity parameters, and enabled options.
5.
Press the CONFIGURATION tab button.
6.
On the Configuration page, in the left folder tree, select the parameter group to change.
7.
Configure the system according to site preferences.
An overview of the configurable system attributes is given in the tables below.
Contact a GE service representative for items marked “Editable by GE Healthcare personnel
only.”
8.
On each page, after making changes, press COMMIT to put the changes into effect.
4.2.1 Site configuration
The following information can be viewed regardless of service mode. However, some fields are
editable only by sites with Advanced Service mode enabled (
Section 4.12 Enabling Advanced
), and some are editable only by GE Healthcare personnel.
Table 4-2 Site tab parameters
Parameter
Definition
Site Name
Name of specific site where system is installed.
Street Address 1
Street address of site where system is installed.
Street Address 2
Street address, continued, of site where system is installed.
City
City where system is installed.
State/Province
State or province where system is installed.
Postal Code
Postal or ZIP code where system is installed.
Department Name
Department name (for example, Radiology) responsible for system
maintenance.
Telephone
Phone number of department responsible for system.
Fax
Fax number of department responsible for system.
Contact Person Name
Person responsible for system maintenance.
Contact Person Phone
Phone number of person responsible for system maintenance.
System ID
Editable by GE Healthcare personnel only.
System identifier assigned by GE Healthcare following system
installation and prior to customer turnover.
This field is a GE proprietary field and cannot be used for any purpose
other than the System ID. Violation of this mandate will result in Insite
and/or other features not working. Customer IT personnel need to use
other means for DICOM filtering on their DICOM machines.
Installation Date
Editable by GE Healthcare personnel only.
Date system installation is completed, created by GE Healthcare.
Service Contract Number
Editable by GE Healthcare personnel only.
Customer’s Service Contract number, created by GE Healthcare.
Contract Expiration
Editable by GE Healthcare personnel only.
Customer’s Service Contract expiration date, created by GE
Healthcare.