GE H
EALTHCARE
O
PTIMA
XR200
AMX
/XR220
AMX
D
IRECTION
5336113-1EN, R
EVISION
10
I
NSTALLATION
M
ANUAL
Chapter 4 Configuration
Page 93
6.
Disconnect the J5 plug from the Spyder board and plug into End C of the main cable from the
PCB Module.
Plug End B of the main cable into Spyder J5.
See
to see how connections appear after re-wiring is complete.
Figure 4-21 Cable connection to Spyder board
7.
Remove the top-left mounting fastener on the Spyder board and mount the ground terminal,
End D, of the main cable underneath the head of the fastener. See
Figure 4-22 Ground connection
End C
Spyder Board
J5 Plug
J5 Connector
End B
End D
Ground Fastener