Chapter 3 Installation
Page 35
1.
Unpack the unit and remove all packaging materials.
2.
Remove and unpack any components shipped in the system package, such as the service
keyboard and external CD/DVD drive.
3.
Visually inspect the system and components for any damage that might have occurred during
shipping.
-
If no damage is observed, continue with the installation.
-
If damage is found, contact GE Service or the installation specialist.
Section 3.3 Replacing the electrical plug
In regions that use an electrical plug other than the style used in the United States, remove the
electrical plug and replace it with the appropriate hospital-grade plug as dictated by local
regulations.
The replacement electrical plug must meet the following electrical specifications:
-
Hospital-grade plug
-
100-240 VAC
-
1440 VA
-
50/60 Hz
To replace the plug:
1.
Remove the three cross-recessed screws that hold the plug together.
2.
Remove the plug.
3.
Remove the cross-recessed screw holding the wires to the plug.
4.
Attach the wires to the new plug as required:
-
Pin 1, 120/220VAC, hot wire, brown
-
Pin 2, 120 neutral (220 hot) wire, blue
-
Pin 3, earth ground wire, green/yellow
5.
Assemble the new plug.
6.
Test the plug operation.
Section 3.4 Turning on the system
The lighted ring around the power button on the console indicates the system’s power state:
•
Green means the system is on.
•
Solid blue means the system is in a power transition state.
•
Blinking blue means the system is in standby.
•
No light means the system is off.