GE H
EALTHCARE
O
PTIMA
XR200
AMX
/XR220
AMX
D
IRECTION
5336113-1EN, R
EVISION
10
I
NSTALLATION
M
ANUAL
Page 64
Section 4.2 Using the System Configuration Tool (SUIF)
4.2.7 Options configuration
The Optima XR200amx/XR220amx systems are delivered with a base set of operating software
that is configured for immediate operation.
Optional software and hardware purchased by the site must be enabled by GE Healthcare service
personnel, such as:
•
Wireless network connectivity
•
DAP (Dose Area Product) meter (standard with Optima XR220amx, optional for Optima
XR200amx)
•
Repeat Reject Analysis (RRA) (Optima XR220amx or Optima XR200amx with digital upgrade
only)
•
Auto Protocol Assist (APA) (Optima XR220amx or Optima XR200amx with digital upgrade
only)
•
Generator Power - 30kW
4.2.8 InSite configuration
Contact a GE service representative to configure InSite ExC settings.
4.2.9 PNF configuration
The Firewall option at the top of the PNF area must be set to On. If it is set to Off, select TURN
FIREWALL ON.
•
When the firewall is off, all network traffic in both directions is allowed. This setting is used
primarily for troubleshooting.
•
When the firewall is on, it allows only the network communications that are specified by its
configuration and rejects all the rest.
After modifying any PNF options, select RESTART FIREWALL to apply the changes.
4.2.9.1 PNF GUI Control area
The PNF GUI Control area, near the top of the screen, contains these options:
Table 4-9 PNF GUI Control options
Option
Action
Filter Settings
Configure filter settings that allow access through the firewall by
adding allowed services and IP addresses, and removing filters.
Backup/Restore
Options:
BACKUP CURRENT FILTERS
: Create a backup of the current filters.
RESTORE FROM BACKUP FILTERS
: Restore the filters to a
previously backed up set.
RESTORE TO FACTORY DEFAULTS
: Reset all filter settings to
those originally installed.
REMOVE BACKUP FILTERS
:
Network Tools
Show the firewall rules that are currently in effect on this system.
Select
REFRESH
to refresh the list of rules.
Configure PNF
Select which red interfaces the filter settings should be applied to.
Select
UPDATE RED INTERFACES
to apply the changes.