GE H
EALTHCARE
O
PTIMA
XR200
AMX
/XR220
AMX
D
IRECTION
5336113-1EN R
EVISION
10
I
NSTALLATION
M
ANUAL
Page 120
>
Insert the flat field phantom in the collimator rails.
>
Set SID to the on-screen instructions.
>
Make an X-ray exposure.
>
When image appears, touch anywhere on the image to return to the previous
screen.
>
Make another X-ray exposure.
>
When image appears, touch anywhere on the image to return to the previous
screen.
>
Remove the flat field phantom, and replace the DAP meter if applicable.
While the test is running, the status bar at the bottom of the screen will display status
messages.
When the tests are complete, the results screen will be displayed.
5.
Scroll to the right side of the screen to see the pass/fail results.
All tests should pass. If there are any failures, contact GE Service.
6.
To quit the QAP session and return to Applications, press EXIT. Or, to return to the previous
QAP screen, press BACKUP.