Section 4.2 Using the System Configuration Tool (SUIF)
4.2.4 Hardware configuration
4.2.5 Generator configuration
Application Language
Language used for application screens.
QAP Timeout (Optima
XR220amx or Optima
XR200amx with digital
upgrade)
Interval between QAP tests, in days. When time-out is reached,
system users see a warning flag indicating that a QAP test should be
performed.
VA Site
Editable by GE Healthcare personnel only.
Applies to USA Veterans Affairs facilities.
Yes or No.
Screen Blank Timeout
Period of inactivity (no user input or processing activity) after which
the screen will go blank, in minutes.
Image Mask Adjustment
(Optima XR220amx or
Optima XR200amx with
digital upgrade)
Adjust the mask that is applied to the acquired image from the
detector. Options: 3mm, 2mm, 1mm, 0, -1mm, -2mm, -3mm.
Table 4-4 General tab parameters
Parameter
Definition
Table 4-5 Hardware tab parameters
Parameter
Definition
Collimator Light Maximum
On Time
Time, in seconds, that the collimator light will remain on.
Collimator Lamp Install
Date
Date the collimator lamp was installed. YYYY-MM-DD format.
Table 4-6 Generator tab parameters
Parameter
Definition
Tube HUR Alarm
Percentage of the tube thermal capacity limit, from 0 to 100. When this
limit is reached, X-ray exposure is inhibited and an alarm sounds.
Tube ID
Tube ID. Not editable.
Tube Serial
Serial number of the tube. Confirm the serial number matches that of
the attached tube.
Tube Install Date and
Time
Date and time the tube was installed. YYYYMMDD HH:MM:SS
format.