Chapter 3 Installation
Page 43
3.5.1.5 Taking ground measurements
1.
Turn off system power and main circuit breaker.
2.
Connect cable A to the grounding pin on the electrical plug.
3.
Connect cable B to the first measurement point listed in
.
4.
On the tester, press the green
START
button.
The orange indicator is displayed, the Test Current display value quickly increases from 0.0 to
the preset output current (25.0 A or 30.0 A), and the timer value decreases from 5.0 seconds
to 0.0 seconds.
If the test passes, the green Pass indicator is displayed. Record the displayed resistance value
in
5.
Press
Stop
.
6.
Move cable B to the next measurement (
and
.
The voltage must not exceed 2.5 V (0.1 ohms).
Table 3-3 Ground measurement test points
Component
Measurement point
Measured
resistance
Tube
Tube endcap screws or HV cable nuts
Collimator
Skin guards
Column
Scroll support rivets
Bin
Screw used for mounting strain relief of wired handswitch cable
Tube latch
Tube latch on top cover