GE H
EALTHCARE
O
PTIMA
XR200
AMX
/XR220
AMX
D
IRECTION
5336113-1EN, R
EVISION
10
I
NSTALLATION
M
ANUAL
Chapter 1 Safety
Page 21
Chapter 1 Safety
Section 1.1 Energy sources
Type of equipment and/or method selected to dissipate or isolate stored energy:
-
Allow three minutes for stored energy to dissipate. Stored energy is indicated by safety
lamps on the generator Aux boxes.
-
After waiting period, verify dissipation with volt meter.
Type of equipment and/or method used to ensure disconnections:
-
One lock and tag for each electrical power source
-
Multiple locking device
Table 1-1
Energy source
Yes/No
Location of energy
isolating
means
Magnitude of energy
Electrical
Yes
AC power plug
120 VAC/60Hz
220 VAC/50Hz
Electrical
Yes
Circuit breaker
156VDC
Electrical (batteries)
Yes
Battery board
connector J1
156VDC
Pneumatic
No
n/a
n/a
Hydraulic
No
n/a
n/a
Gas/water/steam
No
n/a
n/a
Chemical
No
n/a
n/a
Mechanical motion
No
n/a
n/a
Gravity
No
n/a
n/a
Rear suspension spring
No
n/a
n/a
Column spring
Yes
Scroll lock at
top of column
610 N-m (5400 in-lbs)
Thermal
No
n/a
n/a
Stored energy (generator
capacitors)
Yes
Time discharge and
DVM verification
15kW generator: 2200J
30kW generator: 8800J
Air under pressure
No
n/a
n/a
Oil under pressure
No
n/a
n/a
Water under pressure
No
n/a
n/a
Gas under pressure
No
n/a
n/a
Steam
No
n/a
n/a
Other
No
n/a
n/a