Section 7-3 - Performing Front End (FE) Diagnostics
7-3-8-4 Receive Test
(cont’d)
7) Check the test result message and proceed as follows:
•
If one or more channels were identified as bad, trackball to the
Save Report
button and press
Select
, then contact a GE Medical Systems Service Representative.
•
If no channels were identified as bad, mark the channels suspected as being bad as follows:
•
Trackball to each of the channels in the
Display
area and press select.
•
Trackball to the
Mark bad channel
checkbox and press
Select
.
•
Trackball to
Save Report
and press
Select
, and contact a GE Medical Systems service
representative.
8.) Trackball to the
Low Freq Test
button and press
Select
. The receive test is performed in low
frequency mode, and the test results are displayed in the
Report
dialog box, as shown in
Figure 7-
20
on page 7-36: