Chapter 7 Diagnostics/Troubleshooting
7-23
7-3-7 Beamformer (BF) Diagnostic Tests
(cont’d)
Figure 7-12 BF Tab - FE Diagnostics Dialog Box (RFT Configuration)
4.) 3) Proceed as follows: Trackball to one of the following buttons and press
Select
:
-
• Trackball to the
Board Test
button and press
Select
OR
-
• Perform the Digital Test 0 and Digital Test 1 independently, as follows.
5.) Trackball to the
Digital Test 0
button and press
Select
.
6.) When complete, trackball to the
Digital Test 1
button and press
Select
.
7.) In the
BF
tab, verify that the
Comment
area indicates that the system is ready to perform the test.
8.) Trackball to the
ADC
button and press
Select
. The beamformer digital test and ADC bit tests are
activated sequentially. The test results are displayed, per vector, in the
Ultrasound Scope
window,
as shown in
Figure 7-13
below.
Data Flow Map
Test Description Area
Test Area
Comment Area