Section 8-4 - Front End Parts Replacement
4.) Turn the FE Board over, then remove the brackets that secure the TR4 boards in place as shown
in
Figure 8-44
below.
5.) Carefully pull
upwards
and gently remove the suspected faulty TR4 board.
8-4-2-4
TR4 Boards Installation Procedure
1.) 1.) Carefully insert the new TR4 board in position on the FE Board, making sure it is plugged firmly
in place.
2.) 2.) Return the bracket (previously removed) that secures the TR4 boards to its original location
above the TR4 boards.
3.) 3.) Return and tighten the four screws (two on each side) to fasten the bracket in position and
secure the TR4 boards to the Front board main plate - refer to
Figure 8-43
on page 8-46.
4.) 4.) Carefully return the FE Board to its original position in the FE crate, as described in the
Front
End Boards Installation Procedure
on page 8 - 45
.
NOTE:
When you insert a Front End Board into the crate, ensure that the probe connector's
metal frame slides into its tracks.
Figure 8-46 Removing the Securing Brackets
Figure 8-47 Removing the TR4 Board
Securing Bracket
Securing Bracket
TR4 Boards
TR4 Boards