Chapter 8 Replacement Procedures
8-119
8-8-1-4 B&W Printer Installation Procedure
(cont’d)
2.) Connect the Remote cable, the B/W Video IN cable, and the Power cable, as shown in
Figure 8-
102
below.
3.) Place the B/W printer in its correct final position.
4.) Secure the B/W printer by fastening the belt around it, or together with the VCR (if present).
NOTE:
For shelf installation instructions refer to Color Printer installation in
8-8-4-4 "Color Video Printer
Installation - Procedure1" on page 8-135
.
5.) Re-install the Connectors Panels Cover as described in
8-2-5-4 "Connector Panels Cover
Installation Procedure" on page 8-8
.
6) Reconnect the mains AC power cable from the wall outlet.
7) Turn ON the main circuit breaker located on rear of the Vivid 3N.
8) Turn ON the Vivid 3N by pressing the ON/OFF button on the Control Console.
9.) Select the activation key under System Configuration, VCR/ECG Tab, A/B Swap.
By default, Print B key will be B&W Video printer and Print A key will be Color Video printer. For
more information on switching printer settings, see
3-8-9 "VCR/ECG Tab" on page 3-44
.
Figure 8-102 B/W Video Printer Connection Cables
Remote cable:
From: Left Connectors Panel,
Print Trig
To: REMOTE
Video IN Cable:
From: Left Connectors Panel,
B/W Video OUT
To: VIDEO IN
Power cable
From: Right Connectors Panel, AC outlet
To: AC IN