Chapter 7 Diagnostics/Troubleshooting
7-79
7-4-4-5
General Configuration
Select the
General Configuration
tab in the
Extended Keyboard Test
window to set the type of log files
that is generated, for example an error log or no log at all.
7-4-5
Keyboard Diagnostic Test
The keyboard diagnostic test enables you to check the functionality of the alphanumeric keyboard keys.
1) Access the
HW Diagnostics
menu, as described in
7-4-1 "Accessing the Back End Diagnostic
Options" on page 7-71
.
2) Trackball to the
Keyboard
button and press
Select
. The
Keyboard Test
dialog box is displayed, as
shown below:
Figure 7-52 Keyboard Test Dialog Box
3) Follow the on-screen instructions.
4) When you are finished, trackball to the
Close
button and press
Select
.