GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8-143
8-8-4-8-4
Printer Setup
1.) Return to the WINDOW SETUP top menu using the
button (
Figure 8-127
, above).
2.) Press the
button to display the LAYOUT SETUP menu (
Figure 8-126
).
3.) Select PRN by using the
or
button. The PRINTER SETUP top menu is displayed:
4.) In the PRINTER SETUP menu, select FUNCTION using the
or
button.
5.) Press the
button to display the FUNCTION SETUP menu:
6.) In the FUNCTION SETUP menu, adjust the parameters in accordance with those shown in
Table 8-7
below, by selecting the parameters one-by-one using the
or
button; for each selected
parameter, adjust the value using the
or
button.
8-8-4-8-5
Input Setup
1.) Return to the PRINTER SETUP top menu using the
button (
Figure 8-128
, above).
2.) Select IN by using the
or
button. The INPUT SETUP top menu is displayed:
3.) In the INPUT SETUP menu, select INPUT SEL using the
button.
4.) Adjust INPUT SEL to RGB by using the
or
button.
This setting applies to both PAL and NTSC models.
Figure 8-128 Color Video Printer Window Display (PRINTER SETUP top menu displayed)
Figure 8-129 Color Video Printer Window Display (FUNCTION SETUP top menu displayed)
Table 8-7
FUNCTION SETUP Parameters
Parameter
Setting (PAL and NTSC)
AUTO LIVE
ON
RM2 FUNC
C&PRINT
Figure 8-130 Color Video Printer Window Display (INPUT SETUP top menu displayed)
PRINTER SETUP
/c/l/PRN/i/o
FUNCTION SETUP
INPUT SETUP
/c/L/p/IN/o
Содержание FK000040
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS DIRECTION FK091075 REVISION 04 VIVID 3N PRO EXPERT SERVICE MANUAL ii iii ...
Страница 24: ...GE MEDICAL SYSTEMS DIRECTION FK091075 REVISION 04 VIVID 3N PRO EXPERT SERVICE MANUAL xxii ...
Страница 597: ......