Chapter 8 Replacement Procedures
8-121
8-8-2
Sony VCR Replacement Procedure
NOTE:
This section describes the installation and removal procedure of the SONY SVO-9500MDP2 and
SVO-9500MD2.
The SVO-9500MDP2 (PAL) and SVO-9500MD2 (NTSC) are alike, except for the Color Signal system.
The installation procedure is the same for both models.
Figure 8-104 SONY Video Cassette Recorder (VCR)
8-8-2-1
Tools
No special tools needed.
8-8-2-2
Manpower
One person - 30 mins.
8-8-2-3
Preparations
1) Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
Chapter 3 -
Installation
.
2.) Remove the Connectors Panels Cover as described in
8-2-5-3 "Connector Panels Cover Removal
Procedure" on page 8-8
.
3.) Unpack the VCR.
4.) Check that the VCR AC voltage corresponds to that of your local voltage.
5.) Verify that the switches on the VCR are as shown in
Figure 8-105 "VCR Switches Setting" on page
8-122
. If necessary, adjust accordingly.