GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
7-6
Section 7-3 - Performing Front End (FE) Diagnostics
7-3-1
Accessing the Front End Diagnostic Options
•
In 2D-Mode, press
<Ctrl+F11>
on the alphanumeric keyboard. The
Diagnostic
menu is displayed,
as shown in
Figure 7-1 "Diagnostic Menu" on page 7-3
.
NOTE:
It is recommended to commence Front End Diagnostic tests using the Auto Sequence option, as
described below. To perform tests using the manual options, see
7-3-1-3 "Manual Options" on page 7-9
.
7-3-1-1
Auto Sequence
1.) In the
Diagnostic
menu, trackball to the
Auto Sequence
button and press
Select
. The
Automatic
FE Testing
dialog box is displayed, as shown below:
Figure 7-2 Automatic FE Testing
2.) Select the required number of
Loops
and activate the
Enable
checkbox.
3.) Under
On Failure,
activate either the
Stop
or
Continue
checkbox, as required. Stopping on failure
enables you to review a failed test, the moment it fails. Alternatively, you may prefer testing to
continue on failure and later you may review the report.
4.) Under
Execute Sequence,
trackball to the
Start
button and press
Select
.
Содержание FK000040
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS DIRECTION FK091075 REVISION 04 VIVID 3N PRO EXPERT SERVICE MANUAL ii iii ...
Страница 24: ...GE MEDICAL SYSTEMS DIRECTION FK091075 REVISION 04 VIVID 3N PRO EXPERT SERVICE MANUAL xxii ...
Страница 597: ......