GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
7-22
Section 7-3 - Performing Front End (FE) Diagnostics
7-3-7
Beamformer (BF) Diagnostic Tests
The BF diagnostic tests enable you to check the ADC output and perform a digital beamformer test.
NOTE:
Before performing the BF tests, perform the IMP, VME Bus, RFT and FEC tests, as described
beginning on page
7-13
. If the BF test is performed without first performing these tests, and
errors are identified, you must perform the FEC tests, then the RFT test, the VME Bus test, and
the IMP test, to ensure that the BF board is the source of the error.
NOTE:
When replacing the BF, perform calibration as described in
Section 6-8 "Calibration" on page 6-21
.
NOTE:
IMPORTANT - Do not perform BF Calibration after operating in CW or PW scanning modes
since validation will fail. Perform the BF Calibration procedure only after 2D scanning mode.
1) Access the
FE Diagnostics
dialog box, as described in
7-3-1 "Accessing the Front End Diagnostic
Options" on page 7-6
.
2.) In the
FE Diagnostics
dialog box, trackball to the
BF
tab and press
Select
,
OR
3.) Trackball to the
BF
button in the
Data Flow Map
and press
Select
.
The
BF
tab is displayed, as shown in
Figure 7-11
(or
Figure 7-12
) below:
Figure 7-11 BF Tab - FE Diagnostics Dialog Box (RFI Configuration)