Chapter 10 Periodic Maintenance
10-29
10-7-9-4
Data Sheet for Transducer Source Current Leakage
The test passes when all readings measure less than the values shown in
Table 10-15
and
Table 10-16
.
Record all data on the
Vivid 3N Preventative Maintenance Inspection Certificate
(see page
10-31
).
CAUTION: When power to the unit is ON,
never
switch the Polarity and the status of Neutral . Be
sure to turn OFF power to the unit before switching them, using the POLARITY switch and/or the
NEUTRAL switch.
Failure to comply with this warning may cause damage to the unit!
Table 10-21 Typical Data Sheet for Transducer Source Current Leakage Test
Transducer Tested:
Unit Power
Tester
Power
POLARITY
Switch
Tester
GROUND
or NEUTRAL
Switch
Measurement
ON
NORM
OPEN
ON
NORM
CLOSED
ON
REV
OPEN
ON
REV
CLOSED
OFF
NORM
OPEN
OFF
NORM
CLOSED
OFF
REV
OPEN
OFF
REV
CLOSED