Section 2-2 - Console Requirements
2-2-5-4
Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without requiring the use of extension
cords. Other outlets adequate for the external peripherals, medical and test equipment required to
support this unit must also be present and located within 1 m (3.2 ft) of the unit. Electrical installation
must meet all current local, state, and national electrical codes.
2-2-5-5
Mains Power Plug
If the unit arrives without a power plug, or with the wrong plug, contact your GE dealer. When necessary,
the installation engineer will supply the locally-required power plug.
2-2-5-6
Power Stability Requirements
Voltage drop-out
Max 10 ms.
NOTE:
The Vivid 3N Pro/Expert ultrasound unit can be provided with an external UPS system. Contact your
local GE Service Representative for details.
Power Transients
(All applications)
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of transient, including
line frequency, synchronous, asynchronous, or aperiodic transients.