Chapter 8 Replacement Procedures
8-83
8-5-12-4 BEP1 - ECG Module Installation Procedure
(cont’d)
4) Reconnect the power cable and the COM 2 cable to the rear of the new ECG Module.
5) Replace the BEP Front Panel and secure it with the screw that was previously removed.
6) Carefully return the BEP to its original position.
7) Return both BEP side covers as described in
8-5-2-3 "BEP1 - Left and Right Side Cover Removal
Procedure" on page 8-62
.
8) Push the BEP as far to the right as possible; tighten first the
left
lower
bracket, then the
right
lower
bracket, and finally the
upper left
side bracket. Use two screws to fasten each bracket securely.
9) Reconnect all cables at the rear of the BEP.
10) Return the Vivid 3N rear cover to its original position - see
8-2-4 "Rear Cover Replacement
Procedures" on page 8-7
.
11) Return the Vivid 3N side covers, as described in
8-2-2-4 "Side Covers Installation Procedure" on
page 8-4
.
12) Reconnect the mains AC power cable from the wall outlet.
13) Turn ON the main circuit breaker located on rear of the Vivid 3N.
14) Turn ON the Vivid 3N by pressing the ON/OFF button on the Control Console.
Figure 8-70 New ECG Module inside BEP1
BEP Front Panel
ECG Module
Aligned in Correct Position