Chapter 3 Installation
3-61
3-9-5-2 Validating Communication with the Workstation
(cont’d)
5) Press
<Ctrl+S>
to view the status of the spooler, as shown below:
Figure 3-44 Job Spooler
The
Status
column can contain the following options:
•
Hold
:
Images are in the spooler buffer but transmission has not begun as the exam is
active.
•
Active
:
Images are being transmitted over the network.
•
Done
:
Transmission is complete and all the images have been transmitted over the
network.
•
Fail
:
Transmission failed or is incomplete. An error message will be displayed in the
Error
column.