Chapter 3 Installation
3-19
3-5-5
Video Formats Confirmation
The Vivid 3N ultrasound scanner and VIC may be configured to operate with either PAL or NTSC video
systems, as required.
3-5-5-1
Video Format Confirmation
1) Turn ON the system.
2) Press
Config
.
3) From the System Configuration dialog box, select the
VCR/ECG
tab and make sure either
PAL
or
NTSC
is selected.
4) Click the
Technical Support
tab and make sure Frequency is set to either
60Hz
for NTSC, or
50Hz
for PAL.
5) Click
OK
.
Note:
If the video format is not set correctly, contact an authorized GE Service Representative.