Section 8-2 - Cover Replacement Procedures
8-2-14
Rear Handle Replacement Procedure
NOTE:
This procedure describes the replacement procedure for the Vivid 3N Rear Handle (P/N
225392-5)
8-2-14-1
Tools
Use the appropriate Phillips-type screw drivers and a combination wrench, as described in the
procedures below.
8-2-14-2
Preparation
Shut down the Vivid 3N ultrasound unit, as described in
Chapter 3 - Installation
8-2-14-3
Rear Handle Removal Procedure
1.) Carefully remove the monitor, as described in the
Vivid™ 3 Monitor 15" Removal Procedure
on
page 8 - 28 or the
Vivid 3 17" Monitor Removal - Procedure 1
on page 8 - 30 (as appropriate).
2.) Remove the control console lower cover, as described in the
Control Console Bottom Cover (Upper
Section) Removal Procedure
on page 8 - 17.
3.) Remove the six screws from the metal service cover on the base assembly and remove the cover.
4.) Push the VGA connector slightly to allow the control console bottom cover (lower section) to slide
down sufficiently to expose the rear handle securing screws (see
Figure 8-27
below).
5.) Using a combination wrench, loosen and remove the two hexagon-head screws (one on each side)
that secure the rear handle to the upper console cover, as shown in
Figure 8-27
.
6.) Remove the Rear Handle.
Figure 8-26 Vivid 3 Rear Handle