8-72
Section 8-5 - Back End Parts Replacement
8-5-7
SCSI Board Replacement Procedure
CAUTION:
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO
TAKING THE NECESSARY ESD PRECAUTIONS:
1.ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED ESD
CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (TO THE RIGHT OF
THE POWER CONNECTOR).
2.FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
8-5-7-1
Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the SCSI board
replacement procedure.
8-5-7-2
Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
Chapter 3 -
Installation
.
8-5-7-3
SCSI Board Removal Procedure
1) Remove the Vivid 3N rear and left and right side covers - see the
8-2-4-3 "Rear Cover Removal
Procedure" on page 8-7
and the
8-2-2-4 "Side Covers Installation Procedure" on page 8-4
.
2) Disconnect the power cable from the rear of the BEP.
3) Remove the BEP side covers, as described in the appropriate procedure:
8-5-2-3 "BEP1 - Left and Right Side Cover Removal Procedure" on page 8-62
, or
8-5-3-3 "BEP2 -
Side Cover Removal Procedure" on page 8-64
.
4) If applicable, remove the octopus card holder as described in
8-5-3-7 "BEP2 - Octopus Card Holder
Removal Procedure" on page 8-67
.
5) Disconnect the cable connecting the SCSI board to the MO.
Note:
Counting from the top, the SCSI board is positioned in the second PCI slot.
6) Release the securing screw located on the left of the board bracket.
7) Gently remove the SCSI board from the BEP.
8-5-7-4
SCSI Board Installation Procedure
1) Gently insert the SCSI board into its slot.
Note:
Counting from the top, the SCSI board is positioned in the second PCI slot.
2) Fasten the board with the screw on the left of the bracket.
3) Reconnect the cable from SCSI board to the MO.
4.) If applicable, install the octopus card holder as described in
8-5-3-8 "BEP2 - Octopus Card Holder
Installation Procedure" on page 8-67
.
5) Install the BEP side covers, as described in the appropriate procedure:
8-5-2-3 "BEP1 - Left and Right Side Cover Removal Procedure" on page 8-62
, or
8-5-3-4 "BEP2 -
Side Cover Installation Procedure" on page 8-64
.
6) Reconnect the power cable to the rear of the BEP.