GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
Chapter 3 Installation
3-59
3-9-5-1 Validating Communication with the DICOM Server
(cont’d)
5) If the connection was successfully established, trackball to the
Dicom Storage Server
line and
press
Select
, then trackball to the
Check
button and press
Select
.
•
If the connection is established, a green checkmark is displayed.
•
If the connection cannot be established, a red cross is displayed.
The properties of the DICOM storage server configuration are displayed on the right side of the tab,
for reference.
6.) If the connection was successfully established, trackball to the
Dicom Print
line and press
Select
,
then trackball to the
Check
button and press
Select
.
-
If the connection is established, a green checkmark is displayed.
-
If the connection cannot be established, a red cross is displayed.
7) If both connections were established, click
OK
to exit the configuration procedure. If not, review the
entire procedure to locate the error.