Chapter 4 Functional Checks
4-23
4-4-4-5
CFM Operation Tests
Table 4-37 CFM Operation Tests (739L)
Step
Task
Expected Result(s)
1
Select a Cardiac preset and set the 2D Gain
to 40.
2
Activate CFM and set the following CFM
parameters:
•
Gain: Max possible without color noise
•
Frame Rate:2
•
LVR: MIN setting
CFM Artifacts Test
3
Place the 739L probe on the RMI1425A
Phantom (top view A).
Ensure there are no flash strikes or other color artifacts.