Chapter 7 Diagnostics/Troubleshooting
7-25
7-3-7 Beamformer (BF) Diagnostic Tests
(cont’d)
9)
Under
Auto Detect
, click the
Start
button. Checking of the test results commences and the light
adjacent to the button flashes green and continues to flash while
Auto Detect
is in progress. If a fault
is detected, the flashing light changes to red and an error log is displayed listing the channel/s on
which the test failed.
Alternatively, review the information manually, using the scroll arrows on the right of the
Ultrasound
Scope
window to select a vector between 0 and 63. The selected vector is displayed in the
Vector
#
field and its test results are displayed in the
Spectrum
area.
Check the test results for each vector, as follows:
•
Verify that the RMS level displayed in the
RMS
field is above 70.
•
Verify that the amplitude at frequency 2.5 is within a range of -5 to -20.
•
Verify that the amplitude at frequency 5 is below -50.
•
Verify that the amplitude at frequency 7.5 is below -60.
NOTE:
If the RMS level is above or below the specified level, reboot the system, perform calibration
and repeat the test.
10) Proceed as follows:
•
If no problems were identified, proceed to the FB tests, as described in
7-3-8 "Front Board
Assembly (FB) Diagnostic Tests" on page 7-26
.
•
If no problems were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
•
If errors were identified but the results were marginal, perform calibration, as described in
Section 6-8 "Calibration" on page 6-21
, and then repeat the test.
•
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
Chapter 5 -
Components
and Function (Theory)
. If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
11) When finished, trackball to the
Cancel
button and press
Select
. The
FE Diagnostics
dialog box
is
re-displayed.