GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
Chapter 10 Periodic Maintenance
10-27
10-7-9
Probe Current Leakage Test
10-7-9-1
Definition
This test measures the current that would flow to Ground (from any of the probes) through a patient who
is being scanned, and who becomes grounded by touching some other grounded surface.
10-7-9-2
Generic Procedure
Measurements should be made with the Ground OPEN and CLOSED, with power line polarity
Normal
and
Reversed
, and with the unit OFF and ON. For each combination, the probe must be active to find
the worst case condition.
NOTE:
Each probe will have a certain amount of leakage current, depending on its design. Small variations in
probe current leakages are normal from probe to probe. Other variations will result from differences in
line voltage and test lead placement.
Figure 10-5 Set Up for Probe Current Leakage Test
POWER
OUTLET
H (BLACK)
POLARITY REVERSING SWITCH
MOMENTARY
SWITCH
N (WHITE)
G (GREEN)
LEAKAGE TEST
METER
CONSOLE
PROBE