Using Webmin service interface
43
2062973-002
The
Login to Webmin
dialog box displays.
9.
Closing Webmin
1.
You can exit the monitor’s maintenance mode any time by pressing any keypad or
keyboard key, using the Trim Knob, using the touchscreen or clicking a mouse button. The
monitor will automatically restore the original network configuration to the S/5 or MC
Network and return to the normal monitoring mode.
2.
Restore the original network settings in the service PC.
3.
Disconnect the crossover cable from the monitor and from the service PC.
4.
Reconnect the monitor back to the S/5 or MC Network if applicable.
4.4 Remote access to Webmin using a service PC over the
IX Network
NOTE: This connection method is supported only if the monitor is equipped with the Advanced
Interface Board that provides both the IX and the MC connectors and the monitor is connected
to a live IX Network.
1.
Connect a service PC to the IX Network using a standard network cable.
2.
In patient monitor, select
Monitor Setup
>
Service
. The local browser opens and displays
the
Login to Webmin
dialog box.
3.
Record the IX IP address of the patient monitor:
IX IP address: ______________________
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......