Troubleshooting
157
2062973-002
Service Monitor
Activation Failed
Software activation failed.
Reactivate software. If that does not
help, reinstall the software and then
reactivate it.
Setting activation failed.
Reactivate settings.
Module
voltage low
•
One of the supply voltages
for the acquisition
modules is out of
specification.
•
An analog output voltage
is out of specification.
Log in to Webmin and select
Diagnostics > Hardware Stats
to
diagnose the problem. If there is a
faulty board, replace it.
Application
error:
Webmin
The operating system informs
that the Webmin local
browser was terminated
abnormally.
Select
Audio Pause
to acknowledge
the message.
Replace monitor
battery
The full capacity of the battery
is less than or equal to 50%
compared to the full capacity
of a new battery.
Replace the monitor battery with a new
one.
Battery failure
Battery failure.
Replace battery with a new one.
Condition monitor
battery
Battery needs conditioning.
Condition the monitor battery. See
section
10.4.5. Conditioning a battery
Battery
temperature high
Battery temperature error due
to faulty battery or battery
management error.
Replace battery with a new one. If that
does not help, replace the DC/DC
board.
Speaker failure
1. Speaker cable loose.
1. Connect the speaker cable.
2. Speaker failure.
3. Replace the speaker.
4. CPU tone generator or audio
amplifier failure.
5. Replace the CPU board.
Message
Possible causes
Possible solutions
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......