Troubleshooting
163
2062973-002
11.8.3
Incorrect system time
Audible alarms do not work.
Audible alarms are turned off
(See
Alarms Setup > Audible &
Visual.
)
Enable audible alarms.
Alarm volume is low.
Adjust alarm volume
(
Monitor Setup >
Sound Volumes
).
Speaker failure
Replace the speaker unit.
Speaker cable loose or faulty.
Check that the speaker cable is intact and
properly connected.
Tone generator or audio
amplifier failure
Replace CPU board .
Problem
Possible cause
Recommended action
System time is incorrect when
patient monitor is not
connected to network.
CPU timekeeper battery empty
Replace CPU timekeeper battery.
Time not configured properly
Configure date and time, see
System time is incorrect when
patient monitor is connected to
MC Network.
Network device time
synchronization error.
When adding a new device to the
CARESCAPE Network, the existing devices
on the CARESCAPE Network will
synchronize to the new device’s time. To
prevent potential time synchronization
issues, you should set the new device’s
time to be as close as possible to the time
(within one minute or less) used by the
existing GE devices on the CARESCAPE
Network.
System time is incorrect when
patient monitor is connected to
S/5 network.
iCentral time incorrect.
Configure iCentral time master.
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......