CARESCAPE Monitor B650
154
2062973-002
2.
For more detailed information, select
Advanced
tab
.
NOTE: Refer to the Module Frames and Modules Technical Manual for information about PDM
battery.
11.7 Error messages and codes
The following error messages display in the message field if there is a problem with the patient
monitor.
Refer to the patient monitor’s user’s manual for a complete list of system messages. Refer to
Module Frames and Modules Technical Manual for a list of parameter module specific error
messages.
Battery Quality
"OK"
The full capacity of the battery is more than
50% compared to the full capacity of a new
battery.
"Condition"
Battery requires conditioning.
"Replace"
The full capacity of the battery is less than or
equal to 50% compared to the full capacity of a
new battery.
"N/A"
Battery is not connected to the patient monitor.
Field
Description
Remaining capacity (mAh)
Remaining capacity of the battery in mAh.
Full capacity (mAh)
Full capacity of the battery in mAh.
Full capacity compared to new (%)
Full capacity of the battery compared to the
nominal full capacity of a new battery.
Cycle count
The total count of charging and discharging
cycles of the battery.
Voltage (V)
Battery voltage.
Current (mA)
Battery current. Positive when charging,
negative when discharging.
Temperature (°C or °F)
Battery temperature.
Message
Possible causes
Possible solutions
License expired
A trial license has expired.
Enable the license with a new
activation code, if needed.
Configuration
changes. Restart
required
Pending configuration
changes to platform settings
that require patient monitor
restart.
Power cycle the patient monitor.
Network down
No other network device
observed on the MC Network.
Verify that the patient monitor is
connected to an active network.
Field
Status
Description
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......