Configuration
93
2062973-002
6.
Select
Activate
to start the settings activation process.
•
If immediate settings activation was selected:
a.
Wait until the settings activation is completed and the patient monitor has
performed an automatic restart.
b.
Verify that the settings activation was successful and the patient monitor is running
the activated settings.
•
If after next case end / discharge settings activation was selected:
a.
The patient monitor will show a message
Settings activation on next
discharge/after next case end
on the display until the next case end / patient
discharge. The patient monitoring can be continued normally until then.
b.
Settings activation will start automatically after the next case end / patient
discharge, unless this process is canceled by the user.
c.
Wait until the settings activation is completed and the patient monitor has
performed an automatic restart.
d.
Verify that the settings activation was successful and the patient monitor is running
the activated settings.
7.16.4 Canceling pending settings activation
You can cancel pending settings activation anytime while the message
Setting activation
after next case end / Setting activation after next discharge
is shown on the display.
1.
Log in to Webmin.
2.
Select
Configuration
>
Settings
.
3.
In the
Sub-Modules for Settings
menu, select
Activate
.
4.
Select
Cancel
to cancel settings activation.
Select
To
NOTE
Immediately
Activate the new settings
Immediately.
The patient monitor must be in a
case reset / patient discharged state.
After next case end /
discharge
Take the new settings in use
with a delay, after the next
case end / patient discharge.
This option allows you to initiate the
setting activation process with a
delay during an active patient case.
The new settings will activate only
after the next case end / patient
discharge.
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......