CARESCAPE Monitor B650
198
2062973-002
12.3.11 Detaching the Base unit
Disassemble first:
a.
12.3.2. Detaching the Interface Board
b.
12.3.3. Detaching the Frame Side Housing Decorations
c.
12.3.4. Detaching the Top Cover
d.
12.3.5. Disassembling the main unit into Rear and Front units
e.
12.4.1. Detaching the Front Unit Assembly
f.
12.4.2. Detaching the LCD Display Unit
6. Remove the four T10 screws mounting the
AC/DC board to the back plate.
[Torque 1.6 Nm]
Hold the AC/DC board to prevent it from
falling.
Reassemble in reverse order.
- Attach the 4 screws that mount the AC/DC board [Torque 1.6 Nm].
- Ensure the cable clamp and the spring are in place.
- Reattach the housing back cover so that the 4 snaps locks.
- Attach the 1 screw that mounts the housing back cover.
- Reconnect the AC/DC cable and the protective earth cable to the AC/DC board.
- Place the DC/DC cable in the cable holder.
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......