Disassembly and reassembly
191
2062973-002
12.3.5 Disassembling the main unit into Rear and Front units
Disassemble first:
a.
12.3.2. Detaching the Interface Board
b.
12.3.3. Detaching the Frame Side Housing Decorations
c.
12.3.4. Detaching the Top Cover
1. Set the monitor lying on its back and remove
the two screws (T10) from the bottom.
[Torque 0.8 m]
- NOTE: Be careful not to break the PDM
docking mechanism. Consider detaching it
according to section
2. Set the rear side of the Main Unit facing you.
3. Remove the three T10 screws from the left
side. [Torque 0.8 Nm]
4. Remove the two T10 screws from the right
side next to the Module Frame hinges. [Torque
0.8 Nm].
5. Detach the Module Frame back plate and
disconnect the Hinge Flex Board from the
Module Frame connectors,
see section
NOTE: Handle the Hinge Flex Board with care
when disconnected. Do not twist it or bend it
down.
6. Carefully lead the Hinge Flex Board through
the opening, and turn the Module Frame
against the main unit.
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......