CARESCAPE Monitor B650
112
2062973-002
−
S/5 printers are correctly configured.
−
IX printers and printer locations are correctly configured.
−
Remote service is correctly configured.
−
National requirements are correctly configured.
−
Network is correctly configured.
−
Power Line Frequency is correctly configured.
−
MUSE/12SL is correctly configured.
−
WLAN is correctly configured.
3.
Log out Webmin.
8.3.5 Keypad and remote
Perform this test for the integrated keypad, for the optional secondary display with a keypad
and for the optional remote controller.
1.
Press any hard key in the keypad or remote and verify that the selected menu is opened
on the screen or the selected activity is started.
2.
Rotate the
Trim Knob
control in either direction to move from option to option on the
display until you have an active parameter window or main menu item highlighted. Press
the
Trim Knob
control once to select the highlighted option. Verify that the selected
menu is opened on the screen or the selected activity is started.
3.
Select
to close any open menu and return to the main display.
8.3.6 Mouse
Perform this test only if a mouse is connected to the patient monitor.
1.
Move the mouse until the pointer (arrow) is over an active parameter window or a main
menu item you wish to select and click the left mouse button once to select it.
2.
Select
to close the open menu and return to the main display.
8.3.7 Alphanumeric keyboard
Perform this test only if an alphanumeric keyboard is connected to the patient monitor.
1.
Select
Data & Pages > Admit/Discharge
(or
Start / End Case
).
2.
Select the
Patient tab > Edit Name & MRN
.
3.
Press
Enter
to highlight the
Medical Record Number
field.
4.
Type some text into the
Medical Record Number
field using the connected alphanumeric
keyboard. Include some characters that are specific to the chosen keyboard locale to
verify that the keyboard language configuration is correct.
5.
Select
to close the open menu and return to the main display.
8.3.8 Barcode reader
Perform this test only if a barcode reader is connected to the patient monitor.
1.
Select
Data & Pages > Admit/Discharge
(or
Start / End Case
).
2.
Select the
Patient
tab.
3.
Scan a test barcode that is applicable to your system:
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......