CARESCAPE Monitor B650
162
2062973-002
Alphanumeric Keyboard and barcode scanner issues
Speaker / audible alarm issues
Trim Knob (only) is inoperative. Trim Knob cable loose.
Connect Trim Knob encoder cable to the
user interface board.
Trim Knob encoder faulty.
Replace the Trim Knob encoder.
Hard keys or the power
indicator LEDs in the keypad are
inoperative.
Keypad is faulty
Replace keypad.
User interface board is faulty.
Replace the user interface board.
Alarm light does not illuminate
when there is an alarm
condition on (audible alarms
work and alarm message is
visible)
Alarm light cable is loose
Connect the alarm light cable to the user
interface board.
Alarm light board or alarm light
cable is faulty.
Replace the front unit assembly.
Alarm light does not illuminate
during power-on self test.
“Warm start”
It is less than 15 minutes since the last
power-up (warm start) and this normal
operation.
The alarm light illuminates during the
power-on self test only if it is a cold start
(more than 15 minutes from the previous
start-up).
Problem
Possible cause
Recommended action
Wrong character is displayed
when a key is pressed on
keyboard.
The keyboard locale is not
configured correctly.
Configure the keyboard locale correctly,
see
Wrong character is displayed
when a barcode is read.
The keyboard locale is not
configured correctly.
Configure the keyboard locale correctly,
see
The barcode reader’s language
configuration is incorrect.
Refer to the barcode reader manual.
Barcode reader does not read a
multi-field barcode correctly.
(i.e., the information is not
populated correctly to the fields
in the Admit menu).
Barcode reader parser
configuration is incorrect.
Configure the barcode settings, see
.
The barcode reader is
incompatible with the parser
configuration (field lengths, field
types, delimeters, symbologies
etc.).
Check the barcode information content
and compare it to the current parser
configuration.
Problem
Possible cause
Recommended action
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......