7.8 Setting power frequency
WARNING
Incorrect power line frequency setting could adversely affect ECG and EEG
processing.
1.
Log in to Webmin.
2.
Select
Configuration
>
Power Frequency
.
3.
In the
Power Frequency
window, select the applicable power line frequency.
4.
Select
Save
.
The power frequency configuration takes effect immediately.
7.9 Selecting language and locale
Select the language used in clinical user interface and select the keyboard locale setting for the
alphanumeric keyboard and the barcode reader.
1.
Log in to Webmin.
2.
Select
Configuration
>
Language
.
3.
In the
Language
window, select the patient monitor language and keyboard language:
a.
Select the patient monitor language from the drop-down list and select
Save
.
b.
Select the keyboard locale from the drop-down list and select
Save
.
The language takes effect after the patient monitor is restarted. The keyboard locale takes
effect immediately after it is submitted.
7.10 Selecting national requirements
Activate France specific defaults for the ECG HR adjustment range and the reminder beep
behavior.
1.
Log in to Webmin.
2.
Select
Configuration
>
National Requirements
.
3.
In the
National Requirement
window, select the applicable option:
4.
Select
Save
.
The national requirements changes take effect immediately.
7.11 Configuring modules
You can configure some acquisition module specific settings. The settings are saved to the
permanent memory of the related acquisition module and the settings travel with the module
from one patient monitor to another.
These settings are pre-configured at factory, except the
Assets Settings
. You may need to
re-configure them after corrective maintenance, or for administration purposes.
Value
Description
None
Select the normal defaults.
France
Enable the following country specific monitoring:
- Heart Rate high alarm limit maximum 230.
- Reminder beep will sound every 2 minutes when alarms have
been silenced permanently.
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......