Verification procedure for wireless MC Network infrastructure
C- 5 (6)
2062973-002
•
Take a snapshot of the refreshed WLAN Diagnostics screen, for example, by printing
it to a file.
3.
Move the roll cart to the following test point along the walking path and repeat the step 2
at each test point until you have completed the test plan. While moving the roll cart from
one test point to another, at all times, verify that there are no losses in the ECG
waveforms.
NOTE: A momentary, up to 5 seconds waveform loss is normal during roaming. If longer,
or repeating waveform losses occur between test points, make this an extra test point
and report it the same way as observations in step 2.
Summary and Reporting
Include the following documents to the test results:
1.
Print the full disclosure reports from the CIC about the observed waveform loss situations.
2.
Print the snapshots of the WLAN diagnostics screens that you saved into the service PC.
3.
Mark to the printouts the id of the test point.
Review and evaluate the test results together with GE personnel and the hospital IT specialists.
Summarize, if additional testing is needed and/or if the WLAN infrastructure needs to be
changed.
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......