CARESCAPE Monitor B650
134
2062973-002
Configuring the monitor
1.
Configure ECG:
a.
Select
ECG1
to the screen with waveform and adequate priority.
b.
In the
Parameter Setup > ECG > Setup
, select
ECG1 Lead: II.
2.
Configure invasive pressure:
a.
Select
P1
to the screen with waveform and adequate priority.
b.
In the
Parameter Setup > Invasive Pressures > P1
, select:
−
Label: P1.
−
Scale (mmHg): 0-200 mmHg.
−
Parameter Format: Sys/Dia (Mean).
Configuring the simulator
Refer to the simulator documentation for details on how to use and configure the simulator.
1.
Configure ECG:
−
ECG rhythm: a normal sinus rhythm
.
−
Heart rate: 80 bpm
,
−
Amplitude: 1 mV.
2.
Only for modules with invasive pressure: configure invasive pressure:
−
Sensitivity: 5 µV/V/mmHg.
−
InvBP output: “0 mmHg static pressure” or “atmosphere”.
Testing the synchronization connector
1.
Test the ECG analog output signal:
a.
Connect the oscilloscope probe to the ECG analog output signal (grey wire) and analog
ground (blue or white wire).
b.
Adjust the oscilloscope Time and Volts scales:
c.
Verify that the ECG analog output waveform shown on the oscilloscope screen
corresponds to the ECG1 waveform on the monitor screen.
2.
Test the Arterial BP output signal:
NOTE: Perform this arterial BP test only if an E-PSMP, an E-PSMPW, an E-PRETN or an E-
PRESTN module is in use with the patient monitor.
a.
Ensure that the simulator’s InvBP output channel is configured to 0 mmHg static pressure
or atmosphere.
ECG analog output signal: Grey
Analog GND: Blue or White
Probe Type: x10
Time/Division: 200 mS
Volts/Division: 0.5V
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......