CARESCAPE Monitor B650
68
2062973-002
WARNING
EXCESSIVE LEAKAGE CURRENT - Laser printers are UL 60950/IEC 60950
certified equipment, which may not meet the leakage current requirements
of patient care equipment. This equipment must not be located in the
patient environment unless the medical system standard IEC 60601-1-1 is
followed. Do not connect a laser printer to a multiple socket outlet
supplying patient care equipment. The use of multiple socket outlet for a
system will result in an enclosure leakage current equal to the sum of all the
individual earth leakage currents of the system if there is an interruption of
the multiple socket outlet protective earth conductor.
WARNING
Non-medical equipment does not provide the same level of protection
against electrical shock. Do not touch the patient and any part of
non-medical equipment at the same time. Some examples of non-medical
equipment are laser printers and non-medical computers.
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......