Using the OTDR
Setting Up for OTDR Testing
3
3-11
Table 3-1. Settings for OTDR Tests (cont.)
SETUP Tab
Setting
Description
Cable
tab
MANUAL CABLE
SETTINGS
n
BACKSCATTER
When
MANUAL CABLE SETTINGS
is disabled, the
n
and
BACKSCATTER
values
depend on the selected fiber type. When the
MANUAL CABLE SETTINGS
is
enabled, the tester enters default values that you can change manually.
The
n
value is the index of refraction. Index of refraction is used to calculate
length. The default
n
values defined in the fiber types are suitable for most
applications. Minor differences between the default
n
and a fiber's actual
n
usually do not make enough difference in length to fail a fiber.
Increasing
n
decreases measured length.*
BACKSCATTER
is the backscatter coefficient, which indicates the amount of light
the fiber reflects back to the OTDR (using a 1 ns pulse). This value is used to
calculate event reflectance for OTDR tests.
Enter the backscatter coefficient of the fiber under test, if known. This value is
specified by many fiber manufacturers.
*
For singlemode modules, the 30 km range and an
n
value greater than 1.52 cause an error. These settings produce propagation times
that are beyond the tester’s range.
-continued-
Содержание OF-500 OptiFiber
Страница 12: ...OF 500 OptiFiber Technical Reference Handbook x ...
Страница 18: ...OF 500 OptiFiber Technical Reference Handbook xvi ...
Страница 27: ...Getting Acquainted Powering the Tester 1 1 9 ajt20f eps Figure 1 1 Battery Pack Features ...
Страница 29: ...Getting Acquainted Verifying Operation 1 1 11 ajt56f eps Figure 1 2 Removing the Module ...
Страница 46: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28 ...
Страница 95: ...Using the OTDR Running the OTDR Test 3 3 21 ajt33f eps Figure 3 12 Connecting the OTDR to Spooled Cable ...
Страница 133: ...Using the ChannelMap Function Running the Test 4 4 3 ajt55f eps Figure 4 2 ChannelMap Test Connections ...
Страница 136: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6 ...
Страница 148: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 4 ajt61f eps Figure 6 2 Changing the Connector Adapter ...
Страница 192: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 48 ...
Страница 196: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 7 4 ajt03f eps Figure 7 2 Using the Visual Fault Locator ...
Страница 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 11 36 ...
Страница 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 261: ...Loss Test Methods Method A A 2 B B 3 ajt58f eps Figure B 1 Method A A 2 Reference and Test Connections singlemode shown ...
Страница 263: ...B 5 ajt59f eps Figure B 2 Method B A 1 Reference and Test Connections singlemode shown Loss Test Methods Method B A 1 B ...
Страница 265: ...Loss Test Methods Method C A 3 B B 7 ajt60f eps Figure B 3 Method C A 3 Reference and Test Connections singlemode shown ...
Страница 272: ...B 14 OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 274: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2 ...
Страница 282: ...OF 500 OptiFiber Technical Reference Handbook 8 ...