Creating Custom Fiber Types and Test Limits
Creating Custom Test Limits
10
10-7
Table 10-3. Custom Limit Settings for Loss/Length Tests
Setting Description*
Length
Enter a maximum length for the cabling.
Overall Loss
Enter a maximum total loss for all components and fiber in the cabling. The tester uses this value
only if you do not enter values for
Loss per km
,
Loss per Adapter
, and
Loss per Splice
. If you enter
those three values and an
Overall Loss
value, the tester ignores the
Overall Loss
value.
To avoid confusion, delete the
Overall Loss
value to change it to
N/A
if you enter values for
Loss per km
,
Loss per Adapter
, and
Loss per Splice
.
Loss per km
Loss per Connector
Loss per Splice
Enter a maximum value of attenuation per kilometer for the entire length of the cabling.
Enter a maximum loss for one connector.
Enter a maximum loss for one splice.
The tester uses these values only if you enter all three. Otherwise, it uses the
Overall Loss
value.
When you enter all three values, the tester calculates the maximum loss using those values and the
NUMBER OF ADAPTERS
and
NUMBER OF SPLICES
values on the
Cable
tab in Setup. The
Overall
Loss
value is ignored.
To avoid confusion, delete these values to change them to
N/A
if you do not enter all three.
Power
Enter power values specified for the installation. Typically, you test for power only in Far End
Source mode.
*
Minimum values are usually not used, but are available for special situations.
Содержание OF-500 OptiFiber
Страница 12: ...OF 500 OptiFiber Technical Reference Handbook x ...
Страница 18: ...OF 500 OptiFiber Technical Reference Handbook xvi ...
Страница 27: ...Getting Acquainted Powering the Tester 1 1 9 ajt20f eps Figure 1 1 Battery Pack Features ...
Страница 29: ...Getting Acquainted Verifying Operation 1 1 11 ajt56f eps Figure 1 2 Removing the Module ...
Страница 46: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28 ...
Страница 95: ...Using the OTDR Running the OTDR Test 3 3 21 ajt33f eps Figure 3 12 Connecting the OTDR to Spooled Cable ...
Страница 133: ...Using the ChannelMap Function Running the Test 4 4 3 ajt55f eps Figure 4 2 ChannelMap Test Connections ...
Страница 136: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6 ...
Страница 148: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 4 ajt61f eps Figure 6 2 Changing the Connector Adapter ...
Страница 192: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 48 ...
Страница 196: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 7 4 ajt03f eps Figure 7 2 Using the Visual Fault Locator ...
Страница 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 11 36 ...
Страница 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 261: ...Loss Test Methods Method A A 2 B B 3 ajt58f eps Figure B 1 Method A A 2 Reference and Test Connections singlemode shown ...
Страница 263: ...B 5 ajt59f eps Figure B 2 Method B A 1 Reference and Test Connections singlemode shown Loss Test Methods Method B A 1 B ...
Страница 265: ...Loss Test Methods Method C A 3 B B 7 ajt60f eps Figure B 3 Method C A 3 Reference and Test Connections singlemode shown ...
Страница 272: ...B 14 OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 274: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2 ...
Страница 282: ...OF 500 OptiFiber Technical Reference Handbook 8 ...