OF-500 OptiFiber Certifying OTDR
Technical Reference Handbook
6-18
Setting the Reference in Smart Remote Mode
1
Turn on the tester and remote and let them warm
up for 5 minutes.
2
On the main tester:
set the following on
Loss/Length
tab in Setup:
•
Set
REMOTE END SETUP
to
Smart Remote
.
•
Set
THIS UNIT
to
Main.
3
For an OptiFiber smart remote:
set the following on
Loss/Length
tab in Setup:
•
Set
REMOTE END SETUP
to
Smart Remote
.
•
Set
THIS UNIT
to
Remote.
4
Clean the tester's and source's OUTPUT connectors
and the reference test cord connectors.
5
On the main tester:
press
F
; then select
Set
Loss/Length Reference
.
6
Make the connections shown on the
SET REFERENCE
screen and in Figure 6-6; then press
t
.
7
Optional:
From the
TEST SETUP
screen, you can enter
reference test cord lengths to meet TSB-140
reporting requirements:
Use
L
to highlight a reference test cord number,
press
t
; then enter a length. Press
N
when you
are done.
8
Press
C
OK
to leave the
TEST SETUP
screen.
W
Caution
Do not disconnect the reference test cords from
the tester's or remote's outputs after setting
the reference. If you do, you must set the
reference again to ensure valid measurements.
See “About Setting the Reference” on page 6-13 for
more information about referencing.
Содержание OF-500 OptiFiber
Страница 12: ...OF 500 OptiFiber Technical Reference Handbook x ...
Страница 18: ...OF 500 OptiFiber Technical Reference Handbook xvi ...
Страница 27: ...Getting Acquainted Powering the Tester 1 1 9 ajt20f eps Figure 1 1 Battery Pack Features ...
Страница 29: ...Getting Acquainted Verifying Operation 1 1 11 ajt56f eps Figure 1 2 Removing the Module ...
Страница 46: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28 ...
Страница 95: ...Using the OTDR Running the OTDR Test 3 3 21 ajt33f eps Figure 3 12 Connecting the OTDR to Spooled Cable ...
Страница 133: ...Using the ChannelMap Function Running the Test 4 4 3 ajt55f eps Figure 4 2 ChannelMap Test Connections ...
Страница 136: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6 ...
Страница 148: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 4 ajt61f eps Figure 6 2 Changing the Connector Adapter ...
Страница 192: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 48 ...
Страница 196: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 7 4 ajt03f eps Figure 7 2 Using the Visual Fault Locator ...
Страница 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 11 36 ...
Страница 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 261: ...Loss Test Methods Method A A 2 B B 3 ajt58f eps Figure B 1 Method A A 2 Reference and Test Connections singlemode shown ...
Страница 263: ...B 5 ajt59f eps Figure B 2 Method B A 1 Reference and Test Connections singlemode shown Loss Test Methods Method B A 1 B ...
Страница 265: ...Loss Test Methods Method C A 3 B B 7 ajt60f eps Figure B 3 Method C A 3 Reference and Test Connections singlemode shown ...
Страница 272: ...B 14 OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 274: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2 ...
Страница 282: ...OF 500 OptiFiber Technical Reference Handbook 8 ...