Using the Loss/Length Option
Cleaning the Loss/Length Connectors
6
6-5
Cleaning the Loss/Length Connectors
Always clean and inspect fiber connectors before making
connections. Use 99 %-pure isopropyl alcohol and optical-
grade wipes or swabs to clean the connectors as follows:
Cleaning the OUTPUT Connector
1
Touch the tip of a 2.5 mm lint-free foam swab in
alcohol; then touch the swab to an alcohol-soaked
cleaning pad.
2
Push the swab into the adapter; twist it around 3 to 5
times against the endface, then remove and dispose
of the swab.
3
Dry the endface with a dry swab by twisting it around
in the adapter 3 to 5 times.
4
Inspect the connector regularly with a fiber
microscope, such as the Fluke Networks
FiberInspector Video Microscope.
Cleaning the INPUT Connector
Note
Typically, the input connector requires cleaning
only if it has been touched.
1
Remove the connector adapter to expose the
photodiode lens (see Figure 6-2).
2
Use the method described in steps 1 and 2 under
"Cleaning the Output Connector" to dampen a swab
with alcohol.
3
Twist the damp swab around against the lens 3 to 5
times; then twist a dry swab around against the lens 3
to 5 times.
Содержание OF-500 OptiFiber
Страница 12: ...OF 500 OptiFiber Technical Reference Handbook x ...
Страница 18: ...OF 500 OptiFiber Technical Reference Handbook xvi ...
Страница 27: ...Getting Acquainted Powering the Tester 1 1 9 ajt20f eps Figure 1 1 Battery Pack Features ...
Страница 29: ...Getting Acquainted Verifying Operation 1 1 11 ajt56f eps Figure 1 2 Removing the Module ...
Страница 46: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28 ...
Страница 95: ...Using the OTDR Running the OTDR Test 3 3 21 ajt33f eps Figure 3 12 Connecting the OTDR to Spooled Cable ...
Страница 133: ...Using the ChannelMap Function Running the Test 4 4 3 ajt55f eps Figure 4 2 ChannelMap Test Connections ...
Страница 136: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6 ...
Страница 148: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 4 ajt61f eps Figure 6 2 Changing the Connector Adapter ...
Страница 192: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 48 ...
Страница 196: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 7 4 ajt03f eps Figure 7 2 Using the Visual Fault Locator ...
Страница 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 11 36 ...
Страница 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 261: ...Loss Test Methods Method A A 2 B B 3 ajt58f eps Figure B 1 Method A A 2 Reference and Test Connections singlemode shown ...
Страница 263: ...B 5 ajt59f eps Figure B 2 Method B A 1 Reference and Test Connections singlemode shown Loss Test Methods Method B A 1 B ...
Страница 265: ...Loss Test Methods Method C A 3 B B 7 ajt60f eps Figure B 3 Method C A 3 Reference and Test Connections singlemode shown ...
Страница 272: ...B 14 OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 274: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2 ...
Страница 282: ...OF 500 OptiFiber Technical Reference Handbook 8 ...