![Fluke OF-500 OptiFiber Скачать руководство пользователя страница 122](http://html.mh-extra.com/html/fluke/of-500-optifiber/of-500-optifiber_technical-reference-handbook_2309946122.webp)
OF-500 OptiFiber Certifying OTDR
Technical Reference Handbook
3-48
Table 3-5. Manual OTDR Settings (cont.)
Setting
Description and Selection Guidelines
AVERAGING
TIME
(OFTM-573x
modules)
The averaging time sets the number of measurements averaged together to create the final trace. Longer
times reduce noise on the trace which increases dynamic range and accuracy and reveals more details, such
as smaller non-reflective events. Shorter times decrease test time, but produce a noisier trace with less
dynamic range and decreased visibility of small, non-reflective events.
Auto
: Adjusts the test parameters for a typical test time of 15 seconds per wavelength. This setting is used
for
Auto OTDR
mode and is the default for
Manual OTDR
mode.
Auto Test Time
: Adjusts the test parameters for a typical test time of 5 seconds per wavelength. This
setting speeds up testing but provides less accuracy and increased deadzones.
Auto Deadzone
: Adjusts the test parameters to minimize deadzones, which reveals more detail in and
around reflective events. This usually takes longer than testing with
Auto
or
Auto Test Time
selected.
Testing with
Auto
,
Auto Test Time
, or
Auto Deadzone
selected can take up to 3 minutes per wavelength,
depending on the fiber's characteristics.
Manual time selections:
Lets you set the test time to 15 seconds, 30 seconds, 1 minute, or 3 minutes per
wavelength. The tester adjusts the test parameters to meet the selected time.
Содержание OF-500 OptiFiber
Страница 12: ...OF 500 OptiFiber Technical Reference Handbook x ...
Страница 18: ...OF 500 OptiFiber Technical Reference Handbook xvi ...
Страница 27: ...Getting Acquainted Powering the Tester 1 1 9 ajt20f eps Figure 1 1 Battery Pack Features ...
Страница 29: ...Getting Acquainted Verifying Operation 1 1 11 ajt56f eps Figure 1 2 Removing the Module ...
Страница 46: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28 ...
Страница 95: ...Using the OTDR Running the OTDR Test 3 3 21 ajt33f eps Figure 3 12 Connecting the OTDR to Spooled Cable ...
Страница 133: ...Using the ChannelMap Function Running the Test 4 4 3 ajt55f eps Figure 4 2 ChannelMap Test Connections ...
Страница 136: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6 ...
Страница 148: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 4 ajt61f eps Figure 6 2 Changing the Connector Adapter ...
Страница 192: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 48 ...
Страница 196: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 7 4 ajt03f eps Figure 7 2 Using the Visual Fault Locator ...
Страница 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 11 36 ...
Страница 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 261: ...Loss Test Methods Method A A 2 B B 3 ajt58f eps Figure B 1 Method A A 2 Reference and Test Connections singlemode shown ...
Страница 263: ...B 5 ajt59f eps Figure B 2 Method B A 1 Reference and Test Connections singlemode shown Loss Test Methods Method B A 1 B ...
Страница 265: ...Loss Test Methods Method C A 3 B B 7 ajt60f eps Figure B 3 Method C A 3 Reference and Test Connections singlemode shown ...
Страница 272: ...B 14 OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 274: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2 ...
Страница 282: ...OF 500 OptiFiber Technical Reference Handbook 8 ...