OF-500 OptiFiber Certifying OTDR
Technical Reference Handbook
2-18
Tutorial: Running a Loss/Length Test
This section familiarizes you with the basic operation of
the loss/length option by guiding you through a
loss/length test in Smart Remote mode.
Required Equipment
•
One OptiFiber tester set up as follows:
¾
OFTM-5612B or OFTM-5732 module installed
¾
Connector adapter installed on the tester's
INPUT port. Match the connectors on the fiber
under test.
¾
Charged batteries attached or the ac adapters
connected
¾
Job settings configured as described in the
tutorial “Preparing to Save Tests” on page 2-11
•
A second OptiFiber tester set up like the first one, or
a DTX Smart Remote with a fiber module for the
correct wavelength and the correct connector
adapter installed
•
Two lengths of fiber cabling or a length of duplex
cabling to test
•
Two known-good duplex reference test cords of the
same fiber type as the cabling to be tested. See
Figures 2-3 and 2-4 to determine the connector types
you will need.
•
For testing multimode fiber, two mandrels of the
correct size
•
Fiber cleaning supplies
Step 1: Selecting Loss/Length Test Mode
On the
HOME
screen of both testers, verify that
LOSS
LENGTH
appears in the upper-left corner of the screen. If
it does not, press
A
Change Test
then select
Loss/Length
.
Step 2: Selecting a Fiber Type
Do the following on the tester you will use to initiate
tests (the main tester):
2-1
Press
S
; then press
D
Tab
to select the
Cable
tab.
2-2
Use
LM
to highlight
FIBER TYPE
; then press
t
.
2-3
Use
LM
to highlight the type of fiber you will
test; then press
t
to select the fiber type.
Содержание OF-500 OptiFiber
Страница 12: ...OF 500 OptiFiber Technical Reference Handbook x ...
Страница 18: ...OF 500 OptiFiber Technical Reference Handbook xvi ...
Страница 27: ...Getting Acquainted Powering the Tester 1 1 9 ajt20f eps Figure 1 1 Battery Pack Features ...
Страница 29: ...Getting Acquainted Verifying Operation 1 1 11 ajt56f eps Figure 1 2 Removing the Module ...
Страница 46: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28 ...
Страница 95: ...Using the OTDR Running the OTDR Test 3 3 21 ajt33f eps Figure 3 12 Connecting the OTDR to Spooled Cable ...
Страница 133: ...Using the ChannelMap Function Running the Test 4 4 3 ajt55f eps Figure 4 2 ChannelMap Test Connections ...
Страница 136: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6 ...
Страница 148: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 4 ajt61f eps Figure 6 2 Changing the Connector Adapter ...
Страница 192: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 48 ...
Страница 196: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 7 4 ajt03f eps Figure 7 2 Using the Visual Fault Locator ...
Страница 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 11 36 ...
Страница 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 261: ...Loss Test Methods Method A A 2 B B 3 ajt58f eps Figure B 1 Method A A 2 Reference and Test Connections singlemode shown ...
Страница 263: ...B 5 ajt59f eps Figure B 2 Method B A 1 Reference and Test Connections singlemode shown Loss Test Methods Method B A 1 B ...
Страница 265: ...Loss Test Methods Method C A 3 B B 7 ajt60f eps Figure B 3 Method C A 3 Reference and Test Connections singlemode shown ...
Страница 272: ...B 14 OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 274: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2 ...
Страница 282: ...OF 500 OptiFiber Technical Reference Handbook 8 ...