OF-500 OptiFiber Certifying OTDR
Technical Reference Handbook
6-14
•
Anytime you reconnect a reference test cord to the
tester or other source.
•
Anytime the tester warns you that the reference is
out of date.
The tester requires you to set the reference at these
times:
•
Anytime you change the loss/length module in the
main or remote tester.
•
Anytime you start using a different remote tester.
•
Thirty days after the reference was previously set.
The tester warns you if the reference value is outside of
an acceptable range.
See the sections "Testing in Smart Remote Mode",
"Testing in Loopback Mode", and "Testing in Far End
Source Mode" for details on setting the reference for
each mode.
To see the reference information for the current remote
end setup, select
Set Loss/Length Reference
from the
Functions menu; then press
A
View Settings
.
Table 6-2 shows typical reference values.
Table 6-2. Typical Reference Values
Smart Remote and Loopback Modes
Far End Source Mode*
Module Minimum
Nominal
Maximum Minimum Maximum
OFTM-5612B Multimode
-27 dBm
-20 dBm
-17.5 dBm
-28 dBm
-5 dBm
OFTM-5732 Singlemode
-10 dBm
-7 dBm
-4.5 dBm
-15 dBm
-3 dBm
* For Far End Source mode with OFTM-5xx1 and 5xx2 modules, values are determined by the source and should not exceed the maximum
values. See the source's documentation for the source’s values.
Содержание OF-500 OptiFiber
Страница 12: ...OF 500 OptiFiber Technical Reference Handbook x ...
Страница 18: ...OF 500 OptiFiber Technical Reference Handbook xvi ...
Страница 27: ...Getting Acquainted Powering the Tester 1 1 9 ajt20f eps Figure 1 1 Battery Pack Features ...
Страница 29: ...Getting Acquainted Verifying Operation 1 1 11 ajt56f eps Figure 1 2 Removing the Module ...
Страница 46: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28 ...
Страница 95: ...Using the OTDR Running the OTDR Test 3 3 21 ajt33f eps Figure 3 12 Connecting the OTDR to Spooled Cable ...
Страница 133: ...Using the ChannelMap Function Running the Test 4 4 3 ajt55f eps Figure 4 2 ChannelMap Test Connections ...
Страница 136: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6 ...
Страница 148: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 4 ajt61f eps Figure 6 2 Changing the Connector Adapter ...
Страница 192: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 48 ...
Страница 196: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 7 4 ajt03f eps Figure 7 2 Using the Visual Fault Locator ...
Страница 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 11 36 ...
Страница 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 261: ...Loss Test Methods Method A A 2 B B 3 ajt58f eps Figure B 1 Method A A 2 Reference and Test Connections singlemode shown ...
Страница 263: ...B 5 ajt59f eps Figure B 2 Method B A 1 Reference and Test Connections singlemode shown Loss Test Methods Method B A 1 B ...
Страница 265: ...Loss Test Methods Method C A 3 B B 7 ajt60f eps Figure B 3 Method C A 3 Reference and Test Connections singlemode shown ...
Страница 272: ...B 14 OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 274: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2 ...
Страница 282: ...OF 500 OptiFiber Technical Reference Handbook 8 ...