![Fluke OF-500 OptiFiber Скачать руководство пользователя страница 124](http://html.mh-extra.com/html/fluke/of-500-optifiber/of-500-optifiber_technical-reference-handbook_2309946124.webp)
OF-500 OptiFiber Certifying OTDR
Technical Reference Handbook
3-50
Table 3-5. Manual OTDR Settings (cont.)
Setting
Description and Selection Guidelines
LOSS THRESHOLD
(OFTM-573x
modules)
Lets you set the threshold (in decibels) for reporting loss events. Events at or above the threshold are
included in the event table. The range for this setting is 0.01 dB to 1.50 dB inclusive.
For smaller threshold values (0.01 dB to 0.3 dB), the tester reduces noise on the trace by taking more
measurements for averaging or by using wider pulse widths. Therefore, smaller values may increase test
times or deadzones on the trace.
Auto
sets the threshold to the default value of 0.1 dB. To set the threshold to
Auto
, press
A
Set to
Auto
on the
SET LOSS THRESHOLD
screen.
Note
Setting a loss threshold does not guarantee that all events at or above the threshold will be
found. The fiber's characteristics or using a manual
AVERAGING TIME
or
PULSE WIDTH
may also
limit which events are found.
If the loss threshold is set to less than 0.1 dB, the tester may find many false events due to
inherent imperfections in the fiber.
Содержание OF-500 OptiFiber
Страница 12: ...OF 500 OptiFiber Technical Reference Handbook x ...
Страница 18: ...OF 500 OptiFiber Technical Reference Handbook xvi ...
Страница 27: ...Getting Acquainted Powering the Tester 1 1 9 ajt20f eps Figure 1 1 Battery Pack Features ...
Страница 29: ...Getting Acquainted Verifying Operation 1 1 11 ajt56f eps Figure 1 2 Removing the Module ...
Страница 46: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28 ...
Страница 95: ...Using the OTDR Running the OTDR Test 3 3 21 ajt33f eps Figure 3 12 Connecting the OTDR to Spooled Cable ...
Страница 133: ...Using the ChannelMap Function Running the Test 4 4 3 ajt55f eps Figure 4 2 ChannelMap Test Connections ...
Страница 136: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6 ...
Страница 148: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 4 ajt61f eps Figure 6 2 Changing the Connector Adapter ...
Страница 192: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 6 48 ...
Страница 196: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 7 4 ajt03f eps Figure 7 2 Using the Visual Fault Locator ...
Страница 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 11 36 ...
Страница 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 261: ...Loss Test Methods Method A A 2 B B 3 ajt58f eps Figure B 1 Method A A 2 Reference and Test Connections singlemode shown ...
Страница 263: ...B 5 ajt59f eps Figure B 2 Method B A 1 Reference and Test Connections singlemode shown Loss Test Methods Method B A 1 B ...
Страница 265: ...Loss Test Methods Method C A 3 B B 7 ajt60f eps Figure B 3 Method C A 3 Reference and Test Connections singlemode shown ...
Страница 272: ...B 14 OF 500 OptiFiber Certifying OTDR Technical Reference Handbook ...
Страница 274: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2 ...
Страница 282: ...OF 500 OptiFiber Technical Reference Handbook 8 ...