System Operation:
Vacuum Modes
3-6
Purging
During this procedure the specimen chamber is automatically pumped
down to a lower pressure to remove the old gas, then it is flooded with
the new one (selected in the Vacuum / Mode module) to a higher
pressure. This takes place several times, until the old gas is removed
and the chamber is mostly filled with the new gas.
This is applied when the system is:
•
pumping to the LoVac / ESEM mode from the vented status.
•
in the LoVac / ESEM mode and the gas type is changed.
•
in the LoVac / ESEM mode and the
Purge
button is pressed.
There may be certain applications (e.g., working with a sensitive
sample) where the operator needs to change the purging parameters.
They can be set up, started and terminated in the
Preferences…
dialogue /
Low Vacuum
tab (see Chapter 4).
Note:
This procedure can take several minutes, according to Preferences
setting. Wait until Vacuum status indicates
Vacuum
, because
detectors do not start operation till desired pressure is reached.